top of page

Complete viewing required: every step is crucial to obtain results

PHMB: 5 – 60 ppm (mg/l).

Use Tablet Reagents PHMB

Polyhexamethylene Biguanide

20-PHMB

Be sure to clean all objects that have come into contact with the reagent thoroughly with a brush, water and then distilled water, otherwise the measuring equipment may turn blue over time. This method is calibrated for alkalinity values (M) =120 mg/l and calcium hardness values =200 mg/l. Deviating alkalinity values / calcium hardness values can lead to measurement deviations.

Step-1

• Make sure the measuring chambers are clean beforehand.

• Rinse them with the water you are going to test.

• Turn on the Water Tester using the power button.

2.png

Measuring chamber

Power button

Step-3

• Cover the measuring chamber with the light protection cap.

• Press the ZERO button.

Light protection cap

ZERO button

ZERO

Step-5

• Use the up and down buttons to select the desired parameter for measurement.

Up and down buttons

Step-7

• Add the reagent to the chamber you are going to perform the test in.

• Squeeze the pressed tablet directly into the chamber, in this case, into chamber №1.

Do not touch the tablet with your hands!

Add 1 tablet of PHMB

Дизайн без назви.png

Step-9

• Cover the measuring chamber with the light protection cap.

• Press the T1 button to start the countdown.

Light protection cap

T1 button

Step-2

• Use the special syringe to add 10 ml of water to each of the three measuring chambers.

OR

- Fill the cuvette directly from the water source you are testing.

- Insert the Water Leveler into the chamber.

- Excess water will drain via side channels, leaving exactly 10 ml in each chamber.

Special syringe or Water Leveler

Дизайн назви.png
PoolLAB 2.0 water leveler.png
1.png

Step-4

• Press the T1 button to go to the selection menu for the test parameters.

The T1/T2/T3 button functions:

– Single click: add a parameter.

– Double quick click: remove a parameter.

– Long press (count to three): open the parameter list.

Step-6

• After selecting the desired parameter, click the OK button to confirm your selection.

OK button

1.png

Step-8

• Using a special stirring stick, first crush the pressed reagent tablet, and then stir until it is completely dissolved.

Special stirring stick

Step-10

• Press the OK button, wait until the countdown ends and the test results appear.

OK button

If you want to upload the test results to the cloud, press the Back button (which also functions as Zero). The word “Test” will appear on the screen — then press the Bluetooth/Wi-Fi button.